Hantek — DPO7000 UserManual

To Set the Sample System EN signal easy to observe. At this point, the circle shown in the following figure should be obtained. Figure 7.2 XY mode waveform 4. Observing the measurement results in the above figure and based on the phase difference measurement principle diagram (Figure 7.1), it can be concluded that A/B (C/D)=1, that is, the difference angle between the input signals of the two channels is θ=± arcsin1=90 ° 7.1.3 ROLL Mode Select the ROLL mode, in which the waveform scrolls from right to left to refresh the display. ⚫ If the current delay scanning is turned on, the delay scanning will automatically turn off when the "ROLL" time base mode is enabled.

When the "YT" time base mode is reactivated, delayed scanning will be turned back on. ⚫ The following functions cannot be set in ROLL mode: "Adjust horizontal displacement" (this function is available when the oscilloscope is running in "STOP"), "Delay scanning", "Trigger oscilloscope", "Protocol decoding", "Pass/fail testing", "Waveform recording and playback", "Set afterglow time". 7.2 Acquisition Mode Open [Acquire] , click on the Acquisition method menu, select Normal, Average, Peak, High Res, and the default acquisition method is Normal. DPO7000 User manua l Copyright © Qingdao Hantek Electronic Co., LTD 50