To Set the Sample System EN 7.2.1 Normal In this mode, the oscilloscope samples the signal at equal time intervals to reconstruct the waveform. For most waveforms, using this mode can produce the best display effect. 7.2.2 Average In this mode, the oscilloscope averages the waveforms sampled multiple times to reduce random noise on the input signal and improve vertical resolution. The higher the average frequency, the smaller the noise and the higher the vertical resolution, but the displayed waveform also responds slower to waveform changes. After selecting the "average" mode, the sampling system menu will display the average number of times.
Press the Averages key and rotate the multi-functional knob to set it to 2, 4, 8, 16, 32, 64, 128, 256, 512, or 1024. The default is 2. Figure 7.3 Waveforms before Averaging Copyright © Qingdao Hantek Electronic Co., LTD DPO7000 User manual 51