To Set the Sample System EN 7.3 Memory Depth Storage depth refers to the number of waveform points that an oscilloscope can store in a single trigger acquisition. It reflects the storage capacity of the acquisition memory. Figure 7.5 Memory Depth The following equation shows the relations among memory depth, sample rate, and horizontal time base scale:   MDepth = SRate * TScale * HDivs MDepth ——indicates the memory depth. The unit is pts. SRate——indicates the sample rate. The unit is Sa/s. TScale ——indicates the horizontal time base scale. The unit is s/div. HDivs ——indicates the number of grids in the horizontal direction. The unit is div.

Therefore, under the same horizontal time base scale, a higher memory depth can ensure a higher sample rate. Open [Acquire] , click on the Mem Depth menu, and select the storage depth. ⚫ Single channel mode: Storage depth options: Auto, 25K, 250K, 2.5M, 25M, 50M, 100M, 250M, 500M, 2G. ⚫ Dual channel mode: Storage depth options: Auto, 12.5K, 125K, 1.25M, 12.5M, 25M, 50M, 125M, 250M, 1G. ⚫ Full channel mode: Storage depth options: Auto, 6.25K, 62.5K, 625K, 6.25M,12.5M, 25M, 62.5M, 125M, 500M. 7.4 Fine Open [Acquire] , click on the Fine menu, and the label will light up to indicate fine tuning. The default setting is coarse tuning.

Copyright © Qingdao Hantek Electronic Co., LTD DPO7000 User manual 53