Hantek — DPO7000 UserManual

To Trigger the Oscilloscope EN ⚫ Click on the sens box in the menu to pop up the numeric keypad and set the sensitivity value. ⚫ Click the up and down arrow keys on the right to set the sensitivity value. 8.3 Trigger Source Click on the button area [Trig Menu] to enter the Source menu. The available data sources include: CH1~CH4, EXT. External trigger sources can be used to trigger on the 5th channel while collecting data on all 4 channels. Trigger signals (such as external clocks or signals from the circuit to be tested, etc.) will be connected to the EXT trigger source through the external trigger input terminal [EXT In] connector.

8.4 Trigger Mode The following is a schematic diagram of the acquisition memory. To facilitate the understanding of triggering events, the collection memory can be divided into pre triggered buffer and post triggered buffer. Figure 8.1 Schematic Diagram of the Acquisition Memory After starting operation, the oscilloscope will first fill the pre triggered buffer. After filling, the oscilloscope will start searching and triggering; During the search period, the sampled data will continue to be transmitted to the pre triggered buffer (new data will continuously overwrite existing data).

After searching for a trigger, the pre trigger buffer will contain the events that occurred before the trigger. Then, the oscilloscope will fill the trigger buffer and display the data in the acquisition memory. If the collection is initiated through the [Run/Stop] button, the process will be repeated; If the acquisition is initiated Copyright © Qingdao Hantek Electronic Co., LTD DPO7000 User manual 59