U.S. Department of Defense MIL-HDBK-704-3 — NOT MEASUREMENT

Downloaded from https://www.everyspec.com MIL-HDBK-704-3 Stimulation & Monitoring in series on Phase B and Phase C. Normal Operation - Voltage Distortion Spectrum Calibrated Resistive Unit Under Test (See Note 5) Load (UUT) Equipment or CHASSIS (50 Hz to 50 kHz) Figure TAC 106-1 A B C N FIGURE TAC106-1. Normal operation - voltage distortion spectrum (50 Hz to 10 kHz). 2. For 3 phase 4 wire equipment, neutral connection to also be made within 10 cm of UUT input power terminals. For 3 phase 6. CAUTION: Verify suitability of variable frequency power source and coupling transformer for distortion spectrum testing. 7. Coupling Transformer shown is connected in series on Phase A.

Testing is repeated with Coupling Transformer connected See Note 1 3 wire equipment, line-to-neutral measurements to be made with neutral connection made to power supply neutral. 5. Stimulation and Monitoring Equipment are user defined. This functional block shall provide appropriate inputs Spectrum Analyzer (See Note 4) 4. CAUTION: Verify suitability of instrumentation inputs and/or use appropriate attenuation. Ground Plane 1. Voltage measurements shall be made within 10 cm of UUT input power terminals. 3. CAUTION: Verify suitability of power supply NEUTRAL and GROUND connections. XMFR See Note 6 CAUTION See Note 7 50 ! H (stimulations) and monitor UUT outputs (e.g.:

RPM, signals, data, etc.) Frequency Counter (See Note 4) Coupling True RMS Voltmeter (See Note 4) A B C N See Note 3 CAUTION Variable Frequency Adjustable AC 3-Phase Power Supply Power Source GND Notes: 36