HP — 6. OMAS Calibration

Perform the Vacuum Calibration as follows: 1. In the Service Calibrations submenu, scroll to “Vacuum Calibration” and press Enter . 2. When the following message appears on the front panel, you must select whether you would like to continue with the calibration by pressing the Enter key. Press Back or Cancel to exit the calibration. 3. The Printer will set the default nominal and real values of the Vacuum Fan. 4. Once the calibration is completed, OK will be displayed on the Front Panel. 6. OMAS Calibration The purpose of this Service Calibration is to calibrate the OMAS Module.

The calibration process consists of printing a special calibration plot and then scanning it with the printer.The scanned result values are compared with the actual printed values to obtain a PASS/FAIL result. Perform the OMAS Module Calibration as follows: NOTE: In order to perform this Calibration, you should order the Paper Advance Calibration Kit (Part Number Q1273-60296) which contains two cut sheets of HP Universal Instant-dry Gloss Photo media (36-inches wide by 42-inches long). • In the Service Calibrations submenu, scroll to “OMAS Calibration” and press Enter . Service Calibrations Entering the Service Calibrations Menu 167