System Error: 94:08 Problem Description: Color Calibration failed: This errror message is briefly displayed after failing to calibrate. Corrective Action: Try the following: System Error Codes • There are a few types of media which will cause the calibration to fail if used: • Semi-translucent and transparent media (film). • Media that is not completly flat with a Satin finish will cause the light to reflect in different direc- tions. • Perform Color Sensor Test ⇒ • Check the connections of the cables between the Spectrophotometer and the Carriage Assembly.
• Replace Color Sensor ⇒ Page 392 • Replace Carriage PCA ⇒ Page 334 System Error 94:08.1 Problem Description The color calibration failed because the Spectrophotometer measured a value L*less than 100. This error is not displayed on the Front Panel, but is still logged within the warning error list. Corrective Action Try the following: • When this error is displayed there is a high probability that the embedded Spectrophotometer is dirty and needs to be replaced, perform the Spectrophotometer test to confirm this ⇒ page 148 • Check the cable connections between the Carriage PCA and the Spectrophotometer.
• Replace Color Sensor ⇒ Page 392 • Replace Carriage PCA ⇒ Page 334 System Error 97:01 Problem Description The Automatic Paper Advance Calibration failed. Corrective Action Try the following: • There are a few types of media which will cause the calibration to fail if used: • Semi-translucent and transparent media (film) • Low quality 3rd party media will cause the printed dots to diffuse. 76 Chapter 2 System Error Codes