Modulation EN 8.13.4 Modulation Source Press [Mod] > SignalSrc softkey to select Internal, External or Another Channel as the modulation source. internal Source When internal modulation source is selected, press Shape softkey to select Sine, Square, Ramp, Noise or Arb as the modulation source. The default is Sine. ⚫ Sine ⚫ Square: 50% duty cycle ⚫ Ramp: 50% symmetry. ⚫ Noise ⚫ Arb: Sinc, Exp Fall, Haver Sine, Lorentz, Gause, Dual Tone, ECG. Note: Noise can be used as modulating waveform but cannot be used as carrier waveform. external Source When external modulation source is selected, the Frequency and Shape menu in the modulation menu is grayed out and disabled.
The instrument receives the external modulation signal from the [FSK/Trig/Sync/Extmod] connector on the rear panel. The Duty Deviation is controlled by ±4V signal level on this connector. another ch CH1 and CH2 can be used as modulation sources reciprocally. When CH1 is used as the modulated wave, CH2 can be used as the modulation source. And vice versa. When selecting another channel as the modulation source, please turn on the output of that channel. 8.13.5 Modulation Frequency When internal modulation source is selected, press [Mod] > Frequency softkey to set the frequency of the modulating waveform. ⚫ Use the numeric keypad or direction keys and knob to enter the desired frequency value.
⚫ The modulating waveform frequency range is 2mHz to 1MHz, and the default value is 100Hz. Note: This menu will be grayed out and disabled when other modulation sources except the internal modulation source is selected. Copyright © Qingdao Hantek Electronics Co., LTD HDG3000C User manual 57