curr: Overcurrent protection current value, unit A, floating point. chn: Refer to the CURRent command. Description Overcurrent protection current value curr can be set range: 3 channels model: Channel 1: 0.002A-3.25A Channel 2: 0.002A-3.25A Channel 3: 0.002A-5.05A 4 channels model: Channel 1: 0.002A-3.25A Channel 2: 0.002A-3.25A Channel 3: 0.002A-2.05A Channel 4: 0.002A-1.55A Return value The query returns the overcurrent protection current setting value of the query channel, unit A. For example CURRent:PROTection 1.3,(@2) /* Set channel 2 overcurrent protection current to 1.3A */ CURRent:PROTection? (@2) /*Return 1.3*/ VOLTage Subsystem Set Voltage : VOLTage Command format VOLTage VOLTage?
Function description Set or query the output voltage value of the channel. Parameter volt: Voltage value, unit V, floating point. chn: Refer to the CURRent command. Description Voltage value volt can be set range: 3 channels model: Channel 1: 0V-32.1V Channel 2: 0V-32.1V Channel 3: 0V-8.1V 4 channels model: Channel 1: 0V-32.1V Channel 2: 0V-32.1V Channel 3: 0V-8.1V Channel 4: 0V-16.1V 9