Return value The query returns the voltage setting value of the query channel, unit V. For example VOLTage 5.5,(@2) /* Set the voltage of channel 2 to 5.5V */ VOLTage? (@2) /*Return 5.5*/ Overvoltage Protection Switch: VOLTage:PROTection:STATe Command format VOLTage:PROTection:STATe VOLTage:PROTection:STATe? Function description Set or query the channel overvoltage protection switch status. Parameter swi: OFF | ON | 0 | 1. ch: Refer to OUTP command. Description None. Return value The query returns ON or OFF. For example VOLTage:PROTection:STATe ON,(@1,2) /* Open channel 1 and 2 overvoltage protection */ VOLTage:PROTection:STATe? (@1,2) /*Return ON,ON*/ Set the Overvoltage Protection Voltage:
VOLTage:PROTection Command format VOLTage:PROTection VOLTage:PROTection? Function description Set or query the channel overvoltage protection voltage value. Parameter volt: Overvoltage protection voltage value, unit V, floating point. chn: Refer to the CURRent command. Description Overvoltage protection voltage value volt can be set range: 3 channels model: Channel 1: 0V-32.1V Channel 2: 0V-32.1V 10