Drain Valve subtest This subtest opens (default state is closed) and then closes the drain valve at a 2 Hz. rate for a total of 5 cycles. This should allow support engineers an opportunity to hear the valve functioning. The valve must be left in the default state at the end of the subtest. Motor/Propeller subtest This subtest enables the agitation motor for a total of approximately 5 sec- onds, and then disables the motor. The support engineer should be able to see the propeller rotating during this time.
Float Level Sensor Assemblies subtest Since there is currently no manual or automatic way of activating the float level sensor assemblies, this subtest simply checks that the sensors are in their default state for an empty cleaning tank assembly. System Lid Lock subtest This subtest enables (default state is disabled) and then disables the system lid lock at a 2 Hz. rate for a total of 5 cycles. This should allow support engi- neers an opportunity to both see (if the system lid is open) and hear the lock functioning. The lock must be left disabled at the end of this subtest. System Lid Switch subtest This subtest necessarily involves support engineer interaction.
The support engineer must open and close the system lid so that the system can verify that the system lid switch assembly can reach both states. The support engi- neer must do this within a specified time limit (10 seconds) or the test fails. Once both states have been verified, the subtest is complete. In order to help prompt the support engineer to move the system lid, the ‘warning’ LED should be blinked at a 2 Hz. rate for the duration of this subtest. Heater subtest This subtest starts by verifying that the cleaning tank assembly temperature is within normal operating limits (T env-min °C to T env-max °C).
If so, it enables the cleaning tank assembly heater (at ~10% duty cycle) until either the tempera- ture rises T subtest –limit above the subtest’s starting temperature or a time-out occurs – whichever comes first. If the temperature limit is reached first, then the heater is disabled. It is an error if the time limit expires before the temper- ature limit is reached. The heater should be immediately disabled in case of an error. During this subtest, the support engineer is allowed to keep the sys- tem lid open in order to touch the inside of the sink for manual verification.
Therefore, the subtest temperature limit (T subtest-limit ) should be kept There is no delay to wait for the system to cool back down after this subtest, so repeated execution of this subtest will eventually cause a failure due to the starting temperature being above T env-max °C. The heater subtest will fail if run within 30 seconds of powering on the system. 30