EPSON — Ionizer/Foreign Substance Detection Roller

L-6534VW Operation Guide Maintenance D Clean the top surface of the ionizer by wiping the cleaning stick back and forth 2 to 3 times. E Clean the ionizer's needle and the parts around where the needle is attached with a cleaning stick. Clean the ionizer's needle and the parts around where the needle is attached from the opposite direction. F Use a dry cleaning stick to wipe off any alcohol for cleaning the rollers that remains on the parts you cleaned. c Important: Do not leave any parts wet with alcohol for cleaning the rollers. G Close the unwinder roll cover on the press. H Press Finished to exit Periodic Maintenance mode. B:

Unwinder side ionizer A Display the Periodic Maintenance screen on the touch panel, and then switch to Feed section mode. (1) Press Periodic Maintenance from the Maintenance menu. (2) Press Switch Mode , select Feed section , and then press OK . 295