Downloaded from https://www.everyspec.com MIL-STD-1311D METHOD 3211 COLD-GAS CONDUCTION (Thyratron) The anode voltage shall be applied between anode and grid with the grid approximately at ground potential. The anode resistor shall be adjacent to the anode and no connection shall be made to the filament (or cathode). The temperature of all parts of the tube shall be between 15 ° C and 50 ° C. The frequency of the anode supply voltage shall not exceed 150 Hz. At the minimum limit specified for the anode-supply voltage, or less, not more than two flash discharges of any nature shall occur.
Between the minimum and maximum limits specified for the anode-supply voltage, the tube shall conduct the specified current for at least 1 second, or shall conduct three or more flashes of current. METHOD 3211 3211-1 19 April 1968