Downloaded from https://www.everyspec.com MIL-STD-1311D 8. Conditions of test. The operating parameters for power measurements of the TUT and the type of power measurement (peak pulse, average pulse, CW) shall be as specified on the individual TSS. These parameters shall include the following where applicable: Operating voltages. Operating currents. RF input and output match specifications. 1/ RF input drive power specifications. Tuning procedures. Other special test conditions.

In addition, for tests of power as a function of some variable, for example, f, Er, temperature, etc., the test shall be conducted either with all parameters constant except the specified variable or with allowable parameter adjustments for maximum power after each change in the variable, as specified on the TSS. 9. Precautions. 9.1 Appropriate filters may be used to isolate the frequency of interest or to eliminate harmonic powers and spurious responses in broadband or synchrodyne tubes. 9.2 Care shall be taken to insure that minimum mismatches exist in the circuitry between the TUT and the specified load.

_________ 1/ For magnetrons in which the specified coupling includes an impedance transformation section, the load VSWR measured shall not include the VSWR of this section. For tubes in which the specified coupling does not impose any such section, the VSWR shall be measured adjacent to the output coupling of the tube. METHOD 4250A 4250-5 25 March 1970