Downloaded from https://www.everyspec.com MIL-STD-1311D NOISE POWER DENSITY METHOD 4282 (Microwave tubes) 1. Purpose. The purpose of this method is the measurement of noise power density, as a function of frequency, for traveling- wave tubes (or other broadband power tubes). 2. Definitions and discussion. Noise power density is the ratio of (1) total noise power emitted within a defined narrow measurement bandwidth, to (2) the width of this band. It is normally expressed in watts/MHz or dBm/MHz. 2.1 The measurement bandwidth should be narrow compared both to its own center frequency and to the working bandwidth of the TUT.

The center frequency of the measurement bandwidth can then be swept across the working band of the TUT, to obtain the frequency dependence of the noise power density. 2.2 Even though the internally-generated component of the noise power density may be low, the large bandwidth of a microwave tube can result in a considerable total of internally-generated noise, which may extend to frequencies well outside the nominal bandwidth. Care shall be taken that this out-of-band noise power does not vitiate the measurement. 3. Equipment (method A). The equipment required is shown on figure 4282-1. The If amplifier bandwidth is the measurement bandwidth.

The bandwidth of the tunable bandpass filter must be wide enough to have no effect on the measurement bandwidth, but must also have adequate rejection at the image frequency (normally 60 MHz away), and its stop bands must extend beyond the frequency range over which the TUT emits significant noise power. Isolators or attenuators may be required on either side of the tunable filter to improve the matching. The circuit is shown for the normal condition in which the expected noise power from the TUT in the measurement bandwidth is greater than, or not much smaller than, the maximum signal generator power.

If the noise power is much less than the signal generator power, the output connections from the directional coupler C must be interchanged. 3.1 Procedure (method A). The input of the TUT is connected to a noise source or to a termination, as specified. The tunable bandpass filter is tuned to the frequency at which the noise power density is to be measured. The local oscillator is offset from this frequency by the amount of the If (normally 30 MHz). The signal generator is tuned to the center frequency of the bandpass filter, and its output is attenuated to the point where it has no effect on the output power meter.

The system is calibrated for dynamic range by plotting the power output from the If amplifier as a function of signal-generator power. 3.1.1 The gain of the If amplifier is adjusted so that the noise from the TUT gives a convenient deflection on the output meter. This reading is noted. The output of the signal generator is then increased until the output meter reading increases by 3 dB. 3.1.2 The signal generator is then switched to the power meter, and its power output is measured as P 1 . 3.1.3 The procedure is repeated with the tunable bandpass cavity, and the signal generator both retuned to the next frequency, continuing until the required band has been covered.

3.2 Calculation (method A). The noise power density N o at each frequency is then calculated as follows: a. If P 1 is measured in mW (or watts): C anti P N ) ( 1 10 log 1 0 = MHz watts or mW F Where C = coupling factor, dB, of directional coupler labeled C in figure 4282-1; F = bandwidth of If amplifier in MHz. METHOD 4282 4282-1 19 April 1968