Downloaded from https://www.everyspec.com MIL-STD-1311D HEATER OR FILAMENT CURRENT, OPERATING 1. Purpose. To measure heater or filament current on an operating tube. 2. Introduction. Microwave tubes may require measurement of operating heater/filament current while one or more of the following occurs: a. Pulse or radio frequency currents are flowing in the heater circuit. b. Back-bombardment of the emitter by rejected electrons causes overheating, therefore requiring that heater/filament power be reduced (triodes, magnetrons, crossed field amplifiers). c.
Heater/filament current is non-sinusoidal due to the action of a back-heat compensator control as in (b) or because of the non-linear action of a low capacitance (high-leakage reactance) supply transformer. d. Because the anode or collector is often grounded in microwave tubes, the cathode may have high voltage applied to it, relative to ground. 3. Equipment. 3.1 The meter shall measure true rms values of heater current, including harmonics as present. 3.2 The meter shall be by-passed with a capacitance of such a value to prevent the pulse or rf current components from passing through the meter.
3.3 When the cathode is at high voltage with respect to the ground, the meter shall be enclosed in a metal box or cage at cathode potential so that: a. The meter reading is not affected by attraction of the pointer due to the high electric fields, and b. the meter is not damaged by corona currents passing through it. 3.4 Magnetic shielding between meters, or sufficient distance between meters and magnetic fields or materials, shall be such that the accuracy of the meter readings are not affected. 4. Procedure. When the heater/filament input is adjusted to be within the given limits under operating conditions, the stabilized voltage or current shall be measured. 5. Precautions.
5.1 Personnel should take adequate precautions to protect themselves when there is high voltage on the meter or hv cage. 5.2 When a low-capacitance transformer is used, or when an automatic back-heating compensator is used, the meter should measure true total rms values of heater supply fundamental component and all appropriate harmonics content. 5.3 Measurements should be made using the back-heat schedule, if specified. 5.4 Voltage and current measurements shall be corrected for incidental voltage drops and shunt currents due to the meters, in accordance with standard practice. METHOD 4281A METHOD 4281A 4281-1 25 March 1970