Downloaded from https://www.everyspec.com MIL-STD-1311D 1.3 Frequency jitter. Frequency jitter shall be measured using a suitable microwave interferometer or frequency discriminator (see figure 4318-3). The measurement shall be taken at the point in time of maximum jitter within the usable portion of the discriminated pulse waveform or as specified on the TSS. Means shall be taken to minimize the effect of frequency drift which might occur during this measurement. In cases where amplitude jitter in the test modulator causes an increase in jitter due to frequency pushing, that amount of jitter which is over and above the tube frequency jitter of the TUT may be deducted. FIGURE 4318-3.

Frequency jitter output of interferometer or frequency discriminator. 2. The method of obtaining cumulative distribution data and calculating the rms value of pulse-to-pulse jitter shall generally be as follows: Step 1: Using suitable associated test equipment (see 3 below), sample the jitter trace in discrete progressive cumulative levels and determine the percent of pulses exceeding a particular level during a fixed time interval. The time interval shall be prf onds sec 000 ,1 minimum or sufficiently long to satisfy statistical computations. Step 2: Continue sampling the jitter trace in progressive cumulative steps to enable a probability curve to be constructed. Step 3:

Plot an arithmetic probability curve of cumulative percent of pulses versus the particular increment of jitter. Figure 4318-4 illustrates such a Gaussian distribution of time jitter of the leading edge of the RF envelope with respect to the voltage pulse and also the resultant probability curve that is obtained as the jitter trace is sampled at the 50- percent level in a cumulative incremental manner. Step 4: The rms value of jitter, σ m(1), shall represent that jitter increment between the 50 percent and 84.13 percent points on this curve, see σ m(1) on figure 4318-5. Step 5:

If the distribution is Gaussian, the pulse-to-pulse rms value of jitter may be computed from the following formula, using time jitter as the example. )1( 2 m Jr σ = If the cumulative distribution of jitter is not Gaussian, due to periodic or cyclic contributions from either the tube or associated test equipment, statistical methods may be employed to deduct that portion of the jitter measurements resulting from periodic variations (see 5 below). 3. Test equipment for obtaining measurement of cumulative distribution data.

Figure 4318-6 is a simplified block diagram of one possible arrangement for obtaining cumulative distribution data in the measurement of time, amplitude, and frequency jitter. This method employs an electronic detector circuit. However, other methods such as photomultiplier detector circuits are equally satisfactory and may also be used. METHOD 4318 4318-2 19 April 1968