Department of Defense MIL-STD-1311D — MIL-STD-1311D Test Methods for Electron Tubes — DoD Standard Manual PDF

Downloaded from https://www.everyspec.com MIL-STD-1311D (OBSOLESCENT) METHOD 4441 LEAKAGE POWER (Gas-switching tubes) 1. TR tubes. The leakage power shall be measured after a shelf life of 7 days. The tube shall be mounted as a series tee with the rf power in the main line set for the conditions specified on the TSS. The tube shall be tuned to the reference frequency, and the ignitor current set to the given value. The peak leakage power can be obtained from the following equation: pulse leakage of width x prr power Average power leakage Peak = 1.1 A functional block diagram of the system is shown on figure 4441-1. FIGURE 4441-1. Equipment layout for leakage-power measurements for TR tubes.

2. Dual TR tubes. Couplers (3-dB hybrid) are used having a maximum terminated VSWR of 1.10 over the specified frequency range. Each coupler shall split the power evenly to within 0.25 dB and shall have a minimum isolation of 25 dB. 2.1 Leakage power shall be measured using the circuit shown on figure 4441-2. The leakage power shall be within the limits specified when measured in accordance with methods 4446 and 4451. FIGURE 4441-2. Equipment layout for leakage-power measurement for dual TR tubes. METHOD 4441 4441-1 19 April 1968