Downloaded from https://www.everyspec.com MIL-STD-1311D TUNING SUSCEPTANCE METHOD 4482A (Gas-switching tubes) 1. Purpose. The measurement of susceptance at the specified frequency indicates the degree of accuracy with which the ATR tube is tuned to resonance. 2. Equipment. The tube shall be inserted as a series tee in the specified mount between a matched generator and a matched termination. 3. Procedure. Either method A or B may be used to measure and compute susceptance. 3.1 Method A. The susceptance shall be measured by comparing the phase of the standing wave before the tube with that of a reference tube that is known to be resonant at the specified frequency.
The susceptance shall be computed as follows: go g b λ π + = 4 tan 2 ) 2 1( Where: b = B/Yo = Normalized susceptance of the tube. g = G/YO = Normalized conductance of the tube. = Shift of standing wave minimum position from sample to reference tube. λ go = Guide wavelength (same units as for at reference frequency). For small values of susceptance , can be expressed approximately as: go g b λ π + = 2 ) 2 1( 3.2 Method B. An alternative technique involves the determination of the ratio of Pi, the power incident upon the tube, and P1, the power transmitted to the matched termination, with the signal source set at the specified frequency. The susceptance of the tube is then computed as follows:
2 2 2 2 + = K K g g b )1 ( 4 4 ) 2 1( 2 Where: g = Normalized conductance of the tube. Pi = The power incident upon the tube. P1 = The power transmitted through the tube to the matched load at the reference frequency. 1 P Pi K = 4. Precautions. Care shall be exercised if a directional coupler is used for a measurement of Pi. The power reflected by the tube is comparable in magnitude to the incident power, and the directivity of the coupler may not be great enough to prevent this reflected power from affecting the measurement of Pi. METHOD 4482A 4482-1 18 August 1970