Downloaded from https://www.everyspec.com MIL-STD-1311D NORMALIZED CONDUCTANCE METHOD 4484 (Gas-switching tubes) 1. Purpose. To determine the value of a parameter in an ATR tube which is a measure of its impedance at resonance. When this value is below a specified maximum, it assures that there will not be any appreciable loss of received signal energy to the transmitter during the reception period. 2. Test equipment. The tube shall be inserted as a series tee in the specified mount between a matched generator and a matched termination. 3. Test procedure. 3.1 Method A.
The maximum value of VSWR, r o , occurs at the resonant frequency of the tube, at which the susceptance, b, is equal to zero. To determine the maximum value, the VSWR, r, is measured over a narrowband of frequencies around this resonant frequency, and a VSWR/frequency curve obtained. The normalized conductance is then computed as follows: Normalized conductance, O r g 1 1 = 3.2 Method B. An alternative method for determining the normalized conductance consists of measuring the power incident upon the tube and the power transmitted to the termination with the signal source set at the resonant frequency Fo. The normalized conductance is then computed as follows: )1 ( 2 1 = K g Where:
Pi = The power incident upon the tube. P1 = The power transmitted to the termination. 1 P Pi K = 4. Precautions. Care shall be exercised if a directional coupler is used for a measurement of Pi. The power reflected by the tube is comparable in magnitude to the incident power, and the directivity of the coupler may not be great enough to prevent this reflected power from affecting the measurement of Pi. METHOD 4484 4484-1 19 April 1968