Downloaded from https://www.everyspec.com MIL-STD-1311D METHOD 5226 LINE-WIDTH (Cathode-ray tubes) Two line-width determinations shall be made; one determination at the center of the screen (position A in both electrostatic- and magnetic-deflection types) and another determination at a point along the direction of high-frequency scanning, distant from the center of the screen by .375 inch (9.53 mm) of the maximum bulb diameter. For electrostatic types, this second position is referred to as position B; and magnetic types, as position C. 1. Line width (electrostatic deflection).
The scanning pattern specified in method 5221 (light output), shall be set at the light output or screen current specified. The high-frequency scanning shall be applied to the deflecting plates nearest the screen and the amplitude shall be adjusted to give a line length of approximately 90 percent of the maximum tube diameter. The low-frequency scanning amplitude shall be expanded to make the line structure clearly visible and adjustment shall be made for best focus in the center of the pattern.
Unless otherwise specified, each pair of deflection plates shall have the same potential which may be varied to obtain the best compromise between vertical and horizontal line width, and which shall not exceed ± 100 volts difference from the adjacent anode or shield voltage. The pattern shall be compressed until the line structure first disappears or begins to overlap or shows reverse line structure at the center of the screen. The line width at position A is then given by the quotient of the width of the compressed-pattern transverse to the line structure divided by the number of lines which are being scanned, and shall be within the values specified.
The connection of deflection elements to the low- and high-frequency scanning supplies shall then be interchanged and line-width determination repeated without adjustment of focus or astigmatism voltages. Without changing the focus or astigmatism voltages, the line structure shall be determined at the less favorable of the two alternate B positions. Again, the line width is given by the transverse width of the pattern divided by the number of lines. The line width so determined shall be within the values specified. 2. Line width (magnetic deflection). The procedure specified in 1 above for electrostatic types shall be followed in the case of line-width determination at position A.
Without interchange of connections to the deflecting elements, line-width determination shall then be made at the less favorable of the two alternate C positions. Unless otherwise specified, standard focusing coil JEDEC No. 106 and deflection yoke (cf. EIA-256-A, “Deflecting Yokes for Cathode Ray Tubes”), or equivalent, shall be used. The distance (D) from center of focus coil air gap to reference line shall be as specified. The line width so determined shall be within the values specified. METHOD 5226 5226-1 19 April 1968