Downloaded from https://www.everyspec.com MIL-STD-1311D SPOT POSITION METHOD 5231 (Cathode-ray tubes) The tube shall be operated at the test conditions specified. The tube shall be shielded against external magnetic and electrical influences and the control-grid voltage shall be set at a value which will avoid damage for the screen. 1. Spot position (magnetic deflection). With the tube operated as specified above, the distance between the center of the undeflected spot and the center of the screen shall be within the limit(s) specified. In the case of magnetic-focus types, the focusing current shall be zero. In the case of electrostatic-focus types, the spot shall be adjusted for focus. 2.

Spot position (electrostatic deflection). The tube shall be operated as specified above, and adjusted for focus with each of the free deflecting electrodes connected to anode No. 2. The spot shall be within a square, the dimensions of which are within the limit(s) specified on the TSS, the center of which coincides with the geometric center of the tube face, and one side of which is parallel to a trace produced by one set of deflecting electrodes. 3. Spot displacement (leakage). The tube shall be operated as specified above and adjusted for focus with each of the free deflecting electrodes connected to anode No. 2 through a 10-Meg resistor.

The spot displacement produced by the shorting of each resistor in turn shall not exceed the value specified. An equivalent combination of resistance and displacement limit may be used. METHOD 5231 5231-1 19 April 1968