Downloaded from https://www.everyspec.com MIL-STD-1311D UNIFORMITY, OUTPUT SIGNAL AMPLITUDE, ELECTRICAL OUTPUT METHOD 5413 (Cathode-ray charge storage tube) 1. Purpose. The purpose of this test is the measurement of the evenness of output signal amplitude from the TUT. The test procedure measures the spatially fixed, gradual variations in output signal amplitude. 2. Equipment. a. A deflection system shall produce writing and reading rasters in the TUT, each with linear saw-tooth deflections in orthogonal directions. The line scan in the reading raster shall be orthogonal to the line scan in the writing raster. b.

The TUT shall be monitored on a CRT with a raster scan synchronized to the reading raster, and on an oscilloscope with provisions for selecting the video output from any line of the reading raster. c. The output-signal amplifier shall be linear, compensated for the output impedance of the TUT, and have a frequency response that is flat within 1 dB between frequencies (f 1 ) and (f 2 ), to insure accurate reproduction of the line spread function from the TUT. f 1 = the read scan-line repetition rate (Hz) D RJ (MHz) Where: R is 100 cycles/SSRD (arbitrarily set). J is the read scan-line length as a fraction of the SSRD. D is the active read scan-line duration in microseconds. d.

The writing gun control circuits shall produce a square-wave signal of approximately 10 cycles, synchronized to the line scan and shall maintain constant unblanking level during the active portions of the write cycle. 3. Procedure. a. Operating conditions shall be as specified in the TSS. b. The oscilloscope display shall be synchronized to the line scan and shall display any line within the area specified within the TSS. Output signals from outside the area specified in the TSS shall be disregarded. The line select feature of the equipment shall be used to determine which portions of the output are within the area specified in the TSS. c.

Operate the TUT such that the reading time is maximum, or as specified in the TSS. d. Without erasing, write the entire area specified to the saturation level. e. Stop writing. Use the erase procedure specified in the TSS and adjust conditions to obtain a single erase cycle that will cause the fastest erasing portion of the specified area to just be erased. Immediately proceed to 3f. f. Use a pattern of 10 black and 10 white bars, all of approximately equal width, orthogonal to the read line scan, and write with overlapping scan lines until the fastest writing portion of the specified area just reaches saturation level. Stop writing. f 2 = METHOD 5413 5413-1 12 December 1972