Downloaded from https://www.everyspec.com MIL-STD-1311D 3. Procedure. a. The TUT shall be operated in accordance with the conditions specified in the TSS or other document calling for this test method. The area in which the measurement(s) is to be made shall be clearly specified in the TSS. b. For this test, the TUT shall be operated in the linear portion of its input-output transfer characteristics. To insure this condition, a background output signal corresponding to a level approximately halfway between cutoff and saturation shall be established. This may be achieved by uniformly charging the specified area up from cutoff or uniformly charging it down from saturation. c.
The TUT writing gun input signal shall be a constant amplitude level that produces an output signal which falls within the linear portion of the output characteristics. d. In a single cycle, write a pattern of one or more widely spaced parallel lines at the sweep rate specified in the TSS. e. Using the read deflection system described above in 2b, display the output signal on the test oscilloscope. The output pulse shall first be displayed at a convenient height and time base, clearly showing the minimum and maximum levels of modulation. The time base shall then be expanded to clearly display the full output while assuring that the tails of the base still appear. Record the following:
(1) The output pulse waveform and the time base of the test oscilloscope in microseconds. (2) The read scan length as a fraction of the SSRD [symbol (J)]. (3) The active read scan-line duration in microseconds [symbol (D)]. f. Figure 5422-2 shows a representative output waveform that may be expected. Also shown in figure 5422-2 is its relationship relative to the storage surface area. g. Using the data obtained in 3e above, prepare a table of at least 20 amplitude points of the output pulse spaced at regular time intervals. The end points of the pulse must be zero amplitude but it is not necessary to center the pulse on the abscissa.
A convenient time interval between data points (in microseconds), t, shall be selected for this purpose and recorded on the table. h. An example of the data table called for in 3g is shown in table I and consists of the tabulated amplitude values taken with a spacing of t as indicated on figure 5422-2. The recorded time calibration of the table is 0.02 µ s/interval. The values assigned to X shall be consecutive integers such as are illustrated in table 5422-1. i. The MTF is obtained directly by taking the Fourier transform of the output (line spread function) signal. j. A reference for the theoretical basis of MTF and also an aid in the data reduction process, is:
Modern Optical Engineering; The Design of Optical Systems , Warren J. Smith, McGraw-Hill, 1966, p. 308-318. (If the manual data reduction is employed, the above reference provides a detailed step-by-step procedure.) METHOD 5422 5422-2 12 December 1972