Downloaded from https://www.everyspec.com MIL-STD-1311D d. An oscilloscope capable of displaying a single line of the output signal shall be provided and shall have waveform expansion capability either internally or by means of associated equipment. e. A camera or other recording means shall be provided to record the output signal. f. It is suggested (although not necessary for this test) that a picture (e.g., TV) monitor, synchronized to the read scan, also be used to display the video signals. 4. Procedure. a. The TUT shall be operated in accordance with the conditions specified in the TSS. The area in which the measurement(s) is to be made shall be clearly specified. b.

The TUT writing gun input signal shall be a constant amplitude level such that approximately 20 write cycles are needed to charge the storage target from cutoff to saturation. The writing scan lines shall be overlapping. c. Without erasing, write the entire area specified to the saturation level. d. Stop writing. Use the erase procedure specified in the TSS and fully erase the specified area of the TUT. e. Using the deflection system described in 3a and the writing cycle specified in 4b, write with a sufficient number of cycles to produce a measurable output signal on the test oscilloscope. f. Select a single line and display the portion of the video signal from the area specified.

The size of the area selected shall be small enough that shading does not influence the measurement. Record both the output signal at the oscilloscope and the number of writing cycles that were required to write the target to that signal level. g. Without erasing, write a sufficient number of additional cycles to produce an incremental change in output signal. Typically each increment in signal level should be approximately equal to the peak-to-peak disturbance signal. Record the output signal at the oscilloscope and the cumulative number of write cycles that were required to write to this new signal level. NOTE:

For tubes that exhibit short reading time (i.e., such that signal level change during the entire measurement time is significant) or which operate in a destructive readout mode, it is necessary to erase to a fixed level prior to each multiple write cycle. This may be accomplished by using a long erase time or by writing to saturation prior to each erase cycle. h. Repeat 4g a number of times, recording the output signal and the cumulative number of write cycles each time, until the output signal indicates the target has reached saturation. i. An illustration typical of the data obtained above is shown in figure 5424-1.

This is presented in graphical form in 5424- 2, which is a plot of output signal current, I S , with its peak-to-peak target disturbance signal current, I S , as a function of the number of write cycles, N. j. Make a plot of the data obtained in 4f through 4h in the form shown on figure 5424-2. k. To determine dynamic range from this plot, construct a staircase within the boundaries of the two curves, as depicted in figure 5424-3. Observe that figure 5424-3 is the dynamic target transfer characteristic, and that the total number of vertical elements, Lpp, in the staircase is the dynamic range, peak-to-peak disturbance level. This may be expressed in decibels using the formula:

Dynamic range, peak-to-peak disturbance level = 20 log Lpp (dB) NOTE: For many applications, the rms target disturbance level is more meaningful, and is approximately equal to 20 percent of the peak-to-peak target disturbance level. That is, the total number of vertical elements, L rms 5 Lpp, and; Dynamic range, rms disturbance level = 20 log L rms (dB) METHOD 5424 5424-2 12 December 1972