Downloaded from https://www.everyspec.com MIL-STD-1311D BACKGROUND, CONTAMINATION, AND PHOTOSENSITIVITY METHOD 6201 (Radiation counter tubes) 1. Background and contamination. The tube response (count rate) determined using the circuit specified shall be not more than the limit specified. The tube shall be mounted in a horizontal position and may be shielded from extraneous radiation fields by lead of a maximum thickness of 2 inches (50.80 mm), with aluminum of a maximum thickness of .125 inch (3.18 mm) interposed between it and the tube. The counting time and the tube operating voltage for this test shall be as specified. 2. Photosensitivity.

The photosensitivity shall be measured after a period of not less than 60 minutes during which time the tube has not been exposed to ultraviolet light. The tube response (count rate) shall be determined, using the circuit specified. The source of light radiation, the position and orientation of the tube with respect to the light source, the voltage at which the tube is operated, and the duration of the count shall be as specified. The testing procedure shall be as follows: Determine count rate with light off, then determine count rate with light on. The difference between the two count rates is the count rate due to photosensitivity.

The difference shall be not more than the limit(s) specified. 3. Background, contamination, and photosensitivity. The tube response (count rate) due to the combined effects of background, contamination, and photosensitivity determined using the circuit specified, shall be not more than the limit(s) specified. The tube may be shielded from extraneous radiation fields by lead with aluminum interposed between it and the tube, with the maximum thickness as specified in 1 above. METHOD 6201 6201-1 19 April 1968