Downloaded from https://www.everyspec.com MIL-STD-1311C 31 January 2001 3.2.3 Calibration. 3.2.3.1 The film mount, including a series of thicknesses of aluminum, is usually read by a film reader service using densitometer to determine energy and intensity. The densitometer calibration shall be traceable to the National Institute of Standards and Technology. 4. Precautions. 4.1 WHEN OPERATING A TUBE WITH INTEGRAL TUBE SHIELDING, AS DEFINED BY THE TUBE MANUFACTURER, IN PLACE BUT WITHOUT EXTERNAL SHIELDING, THE X-RADIATION MAY BE VERY GREAT AND HAZARDOUS.

INTEGRAL TUBE SHIELDING NOT ONLY INCLUDES PERMANENT TYPE SHIELDING BUT ALSO INCLUDES THOSE SPECIFIC SHIELDS RECOMMENDED AND/OR SUPPLIED BY THE TUBE MANUFACTURER FOR OPERATION OF THE TUBE. ADEQUATE PRECAUTIONS SHALL BE TAKEN TO PROTECT PERSONNEL FROM THESE HAZARDS. WHEN THESE CONDITIONS OCCUR, FILM SURVEY TECHNIQUES ARE RECOMMENDED IN PLACE OF THE SURVEY METER. 4.2 Film badges should be assigned to all personnel involved in making x-radiation measurements. 4.3 Before a reading is recorded, precautions should be taken to assure that the indication on the survey meter is a steady one, due only to the tube being measured.

4.4 Survey instrument calibration shall be obtained against known radiation of the same spectral quality as that being measured, or errors as large as 10:1 from the true energy level can be introduced. METHOD 4006 13 September 1972 4006-2