Department of Defense MIL-STD-1311C — MIL-STD-1311C Test Methods for Electron Tubes – DoD Standard PDF

Downloaded from https://www.everyspec.com MIL-STD-1311C 31 January 2001 5. TSS requirements. In addition to the operating conditions for the TUT, the TSS shall specify the following: a. The use of method A or method B. b. If magnetic focus: Focus coil number; Focus coil position; Focus current range. Limits of position of the undeflected spot as: Limiting radius; Position of circle center. c. If electrostatic deflection, the limit(s) of the undeflected spot position shall be stated as a rectangular width, height, and center position. Orientation of the undeflected spot shall be specified in terms of one dimension being parallel to a specified pair of deflecting electrodes.

METHOD 5404A 12 December 1972 5404-2