UNI-T UTG2000B Series — 3.2.3 Frequency Sweep Mode

Figure 3 - 33 Set frequency sweep parameters When starting frequency from low frequency to high frequency. When starting frequency> stop frequency, function / arbitrary waveform generator sweeps from high frequency to low frequency. When starting frequency= stop frequency, function / arbitrary waveform generator outputs fixed frequency. The synchronous signal of frequency sweep mode is low from starting point to midpoint of frequency sweep time, and high from midpoint to end of frequency sweep time. By default, starting frequency is 1kHz and stop frequency is 2kHz, but the range of starting and stop frequency can vary with frequency sweep waveform.

See the table below for the frequency range of frequency sweep wave: Table 3 - 4 3.2.3 Frequency Sweep Mode For linear frequency sweep, waveform generator changes output frequency in a linear way during frequency sweep; for logarithmic frequency sweep, waveform generator changes output frequency in a logarithmic way. Linear frequency sweep mode is default.To change it, please press soft key Type Log on the interface for starting frequency sweep mode (please press Return first to enter the interface for selecting frequency sweep waveform).

UTG2000B Series 93 Carrier waveform Frequency UTG2122B UTG2082B UTG2062B Min Max Min Max Min Max Sine 1µHz 120MHz 1µHz 80MHz 1µHz 60MHz Square 1µHz 30MHz 1µHz 25MHz 1µHz 25MHz Ramp 1µHz 5MHz 1µHz 4MHz 1µHz 3MHz Pulse 1µHz 30MHz 1µHz 25MHz 1µHz 25MHz Arbitrary 1µHz 25MHz 1µHz 20MHz 1µHz 15MHz