Figure 3 - 35 Select logarithmic frequency sweep 3.2.4 Frequency Sweep Time Set the time from starting frequency to stop frequency, which is 1s by default and in the range of 1ms~500s. To change it, you can use multi-functional knob and direction key on the interface for selecting frequency sweep mode or press soft function key Param SweepTime , input number through numeric keyboard and press corresponding soft key of unit. Figure 3 - 36 Set frequency sweep time 3.2.5 Select Trigger Source The signal generator generates frequency sweep output upon receiving a trigger signal and then waits for the next trigger signal. The trigger source of frequency sweep can be internal, external or manual.
To change it, you can use multi-functional knob and direction key on the interface for selecting frequency sweep mode or press soft function key Param TrigSrc 94 UTG2000B Series