UTD2000 Series User Manual 4.5 Explanation of nouns (1) Trigger source : trigger can be achieved from several signal sources: input channels (CH1 and CH2), external triggers (EXT and EXT/5) and AC line. External trigger (EXT TRIG): meaning to set the triggering source base on given external connection to the “EXT TRIG” input channel. This can be signification application for UTD2000 series to perform data acquisition using CH1 and CH2 based on an external clock signal connects to the “EXT TRIG” input channel. External trigger signal connects to “EXT TRIG” input terminal is used for EXT trigger source selection.
When selecting EXT as trigger source, external trigger levels voltage range limited to -3V and +3V. AC line trigger: use AC line power as trigger source. This can be signification application to observe the signals correlated to AC line power for applications in lighting equipment and power supply equipment R&D development. (2) Trigger mode :deciding the behavioral pattern of DSO under no trigger event conditions. Three trigger modes are provided for DSO: auto, normal and single. Auto trigger: Set to automatic trigger mode. The DSO will continuously perform data acquisition without triggering signal. When there is trigger signal, DSO will automatically synchronizing with the trigger signal.
Attention: when scanning waveform is set as 50ms/div or slower time base, no trigger signal is allowed under “auto trigger” mode. Normal trigger: Set to normal trigger mode. The DSO will only perform data acquisition only when trigger conditions are satisfied. When trigger condition is not satisfied, DSO stops data collection. Single trigger: Set to single trigger mode. The DSO will only perform one cycle of data acquisition only when trigger conditions are satisfied. When trigger condition is not satisfied, DSO stops data collection. 30 without triggering signal. Normal Set to normal trigger. The DSO will only perform data acquisition when there is triggering signal.
Single Set to single trigger. The DSO will only perform 1 cycle of the data acquisition when there is triggering signal. Coupling AC Block the DC component of triggering signal DC Pass through DC and AC components of triggering signa HF Reject high-frequency component of the triggering signal (over 80kHz signals) LF Reject low-frequency component of the triggering signal (below 80kHz signals)