UTD2000 Series User Manual Chapter V Acquire System User can change the data acquisition setting of the DSO by pressing the ACQUIRE button as shown in the figure below UTD2000 series DSO. Figure 5-1 Function Key of Acquire System While ACQUIRE button is pressed, the function menu will change to set the data acquisition setting for the DSO. Table 5-1 Acquire Function Menu By changing the acquisition setting of the DSO, user can observe the input signal at different manner. Example below is a noise waveform acquired by different acquisition setting of the UTD2000 series DSO. Figure 5-1 shows the original signal using sampling acquisition, user can see that the waveform display is very noise.
Figure 5-2 shows the same signal with average acquisition of a 32 number, please observe that the waveform display become much more smooth. Figure 5-1 using sampling method Figure 5-2 using average acquisition of 32 number 31 Function Menu Setting Description Acquisition Mode Sampling Set to sampling acquisition mode. Peak value Set to peak value acquisition mode. Average Set to average acquisition mode. Averages 2~256 At the average sampling mode, set average number by using the multipurpose knob (Average number can set to 2,4,8,16,32,64,128,256) Sampling Real-time Set to Real-time sampling. Equivalence Set to Equivalent sampling. Fast Acq ON Turn on fast acquiring.
Waveform capturing rage up to at 2k waveform/second. OFF Turn off fast acquiring. ——