UTD2000 Series User Manual Figure 12-4 Reducing random noise of signals Steps : (2) Connect the signal to ensure stable display of waveform. See the previous illustration for operation instruction. See the previous chapter for guidance on adjusting the horizontal time base and vertical range. (1) As in the previous illustration, set the attenuation factor of the probe and Ch1. (3) Improving trigger by setting trigger coupling. ① Press TRIG MENU in the trigger zone to display the trigger setup menu. ② Set trigger coupling to Low Frequency Suppession or High Frequency Suppession. By selecting low frequency suppression a high-pass filter is set up.

It filters low frequency signal quantities under 80kHz and allows high frequency signal quantities to pass through. If you select high frequency suppression a low- pass filter is set up. It filters high frequency signal quantities higher than 80kHz and allows low frequency signal quantities to pass through. By setting Low Frequency Suppession or High Frequency Suppession, you can suppress low frequency or high frequency noise respectively and achieve a stable trigger. (4) Reducing display noise by setting the sampling mode.

① If the signal being measured is stacked with random noise and the waveform is too coarse as a result, you can use the average sampling mode to eliminate random noise display and reduce the size of waveform for easy observation and measurement. After getting the mean, random noise is reduced and details of the signal are clearer. Follow the steps below: Press ACQUIRE in the menu zone of the front panel to display the sampling setup menu. Press menu operation key F1 to set acquisition mode to AVERAGE, then press the multifunction control knob to adjust the average number of times in multiples of 2, i.e.

2 to 256, until you get the desired waveform display that meets observation and measurement requirements. (See the figure below) 51