UNI-T UTD2000 Series — UTD2000 Series User Manual

UTD2000 Series User Manual Figure 12-5 Signal noise suppressed ② You can also lower the waveform brightness to reduce display noise Caution :In the average sampling mode the waveform display will update at a slower rate. This is normal. Example 5: Using the cursors for measurement Your oscilloscope can measure 28 waveform parameters automatically. All auto parameters can be measured with the cursors. By using the cursors, you can quickly measure the time and voltage of a waveform. Measuring one step voltage of the step signal To measure one step voltage of the step signal, follow the steps below: 1. Press CURSOR to display the cursor measurement menu. 2.

Press menu operation key F1 to set cursor type to VOLTAGE. 3. Turn the multifunction control knob to set cursor 1 at one step voltage of the step signal. 4. Press SELECT to select the cursor, then turn the multifunction control knob again to set cursor 2 at another step voltage of the step signal. The cursor menu will automatically display the V value, i.e. the voltage difference of that point. See the figure below. Note : When using the cursors to measure time, follow step 2 only and set the cursor type to time. 52