UTD2000 Series User Manual Chapter XIV Technical Information Unless otherwise specified, all technical specifications are applicable to the probes with attenuation switch set as 10Ч as well as UTD2000 series DSO. DSO must first meet the following two conditions to meet those specification standards: The instrument must continuously operate for over half an hour with in the operating temperature. If the operating temperature has exceeds 5 , please perform the “Self-Adjustment ”function in the UTILITY system function.

Any of the specification with “Typical” value represented is for reference only Technical Specification Vertical 56 Acquire System Specification Model UTD2025CL UTD2052CL UTD2102e UTD2052CEX UTD2072CEX UTD2102CEX Real-time 250MS/s 500MS/s 500MS/s 1GS/s Equivalence —— 25GS/s 25GS/s 50GS/s Average N can be chosen among 2, 4, 8, 16, 32, 64, 128 and 256; N is the number of average samples.

Input Channel Specifications Input Coupling DC, AC and GND Input impedance (1MΩ± 1%)//(24pF± 3 pF) Probe attenuation coefficient 1×, 10×, 100×, 1000× Maximum input voltage 400V (DC+AC peak value, 1MΩ input impedance) Horizontal System Specification Scanning scope 5ns/div-50s/div Waveform interpolation Sin(x)/x ampling rate and delay time accuracy ±50ppm (any time interval≥1m) Record length 2×512k sampling point Storage depth Single channel: 25k;Double channel: 12.5k Sampling rate and delay time accuracy ±50ppm(any time interval ≥1ms ) Measurement accuracy of time interval (△T) (full bandwidth) Single time: ± (1 sampling time interval+50ppm×reading+0.6ns) >16 average values:

± (sampling time interval+50ppm×reading+0.4ns)