T3100(1)

Note: Contents of this document are subject to change without notice. 1. Probe Rod 2. Probe Tip 3. Ground Lead 4.Tip Locating Sleeve 5. Adjustment Tool Part Exposition : T3000 Series Probe Assembly Drawing 2 3 5 4 T3100 100MHz 16pF/105pF 100M MAX:2000VDC 1 Instructions T3100 100MHz 100:1 Passive Probe

Стр. 1

T3100(2)

Humidity ≤ 85% (Relative Humidity ) Temperature Operating -10 ℃ --+50 ℃ Non operating -20 ℃ --+75 ℃ Cable Length 120cm Net Weight Maximum Working Input Voltage System Bandwidth Range Compensation Capacitance Input Input Resistance 1:100 Attenuation T3100 Item These characteristics apply to a T3000 series probe installed on a specified oscilloscope.When used with another instrument, the oscilloscope must have an input impedance of 1 M . The instrument must have a warm-up period of at least 20 minutes and be in an environment that does not exceed the limits. Specifications X100: DC~100MHz X100:

3.5pF~10.5pF L 10pF~35pF Ω 100M 0 10MHz 1MHz 100MHz 1 00 0V 1500V 2000V 2500V Maximum Working Voltage Derating Curve (VDC+Peak AC) Adjust trimmer L until seeing flat-top square wave on the display. Low-Frequency probe Compensation Before taking any measurements using a probe,first check the compensation of the probe and adjust it to match the channel inputs.Most oscilloscopes have a square wave reference signal available at a terminal on the front panel used to compensatethe probe.Connect the probe to the signal source to display a1KHz test signal on your oscilloscope. Maintenance 100X 100X