Delta Electronics — 8.2.4 Control Registers

AS Series Module Manual _8 8-8 CR# Name Description Att. Default the lower range is read as the upper range and vice versa. 74 Zero point tracking time for CH2 Range: 5–500 Unit: 100 ms R/W 10 75 Zero point tracking range for CH2 Floating-point format Range: 0–10000; 0: disabled R/W 0 76 77 Calibration points for CH2 Range: 2–20 R/W 2 78–117 Calibrated weight for CH2 Floating-point format Calibrated weight of the calibration points 1–20 R/W - 118 Decimal place for CH1 Range: 0–4 R/W 1 119 Decimal place for CH2 Range: 0–4 R/W 1 120 Alarm 0: warning 1: alarm Bit0: error in the power supply Bit1: error in the module hardware Bit2:

error in the driver board R/W 1 200 State register Refer to the explanation below. R/W - 201 Command set Refer to the explanation below. W 0 210 The maximum peak value for CH1 Floating-point format Maximum peak value for CH1 R - 211 - 212 The maximum peak value for CH2 Floating-point format Maximum peak value for CH2 R - 213 - 214 The minimum peak value for CH1 Floating-point format Minimum peak value for CH1 R - 215 - 216 The minimum peak value for CH2 Floating-point format Minimum peak value for CH2 R - 217 - 222 The time to record for CH1 Unit: 1 ms Range: