heidenhain PWT 100, PWT 101 — Bar graphs for tolerance ranges

8 Encoder diagnostics | Diagnosis for encoders with TTL/HTL/HTLs/ interface Signal deviation in oscilloscope display The table below explains the signal amplitude in the oscilloscope displays. The oscilloscope display is not a function of the product and is intended only as an explanation. Depiction Description Optimum signal deviation Signal deviation at the tolerance limit Signal deviation too large Reference mark position The reference mark signal is at a specified nominal position. In the bar graph of the reference mark position, the position of the black bar indicates the deviation from the optimum position. Figure 46:

Bar graph of reference mark position If certain limits are reached during the signal analysis of the encoder, for example an excessive signal frequency, the product cannot perform the signal analysis correctly. In this case, the drag indicators change and a warning symbol is shown. The drag indicators are set to the maximum values. After the next valid measurement, the values measured so far for the drag indicators are displayed. After a certain time interval (15 s) the display is deactivated (gray). As soon as the device detects the next reference mark, the display is reactivated. HEIDENHAIN | PWT 100, PWT 101 | Operating Instructions | 07/2024 105