heidenhain PWT 100, PWT 101 — Diagnostic values of the PWT display

Encoder diagnostics | Diagnosis for encoders with TTL/HTL/HTLs/ interface 8 Zero crossovers of reference-mark signal In the bar graph of the reference-mark zero crossovers, the positions of two black bars show the deviation of the zero crossovers of the reference mark signal from the specified values. Figure 47: Bar graph of zero crossovers of reference-mark signal After a certain time interval (15 s) the display is deactivated (gray). As soon as the device detects the next reference mark, the display is reactivated. Resetting the drag indicators and warnings You can reset the drag indicators and warnings displayed in the PWT display screen.

Tap the Refresh button The displayed drag indicators and warnings are reset. Diagnostic values of the PWT display Count value The Counts screen (4) shows incremental or absolute values, depending on the selection. When switching to the screen, the counting value is incremental; the display will be set accordingly. If you select Rotatory or Linear in the Display settings screen, the count value display is set to Angle [degrees] or Position [µm] . If you select the option Steps in the Display settings screen, the count value display is set to Steps . If you select a setting other than Off from the settings for Reset the count value with the reference mark:

in the Display options for count value screen, the text Traverse the reference mark will be displayed in the Counts screen. When referencing has been completed, the count value will be shown as an absolute value. Count value Description RM not filled The encoder reference run has not been performed or the option Off has been selected RM filled The encoder reference run has been completed Tapping the display opens the Display settings screen. Further information: "Display options for count value screen ", Page 111, "Display settings screen ", Page 114 106 HEIDENHAIN | PWT 100, PWT 101 | Operating Instructions | 07/2024