8 Encoder diagnostics | Diagnostics for encoders with 1 V PP /11 µA PP /1 V PP + Z1 interface Diagnostic values of the PWT display Count value The Counts screen (4) shows incremental or absolute values, depending on the selection. When switching to the screen, the counting value is incremental; the display will be set accordingly. If you select Rotatory or Linear in the Display settings screen, the count value display is set to Angle [degrees] or Position [µm] . If you select the option Steps in the Display settings screen, the count value display is set to Steps . If you select a setting other than Off from the settings for Reset the count value with the reference mark:
in the Display options for count value screen, the text Traverse the reference mark will be displayed in the Counts screen. When referencing has been completed, the count value will be shown as an absolute value. Count value Description RM not filled The encoder reference run has not been performed or the option Off has been selected RM filled The encoder reference run has been completed Tapping the display opens the Display settings screen. Further information: "Display options for count value screen ", Page 111, "Display settings screen ", Page 114 Reference mark RM (5) shows the status of reference-mark detection. The product uses the following colors for displaying the status:
Color Status Description Blue Reference mark detect- ed The display is activated for approx. 0.5 s. If the reference marks follow in too quick a succession, the display may seem to be continuously active. Gray Reference mark not detected Reference mark not yet traversed or reference mark not detected If the reference mark is not detected correctly during signal evaluation, a signal evaluation warning is additionally displayed next to the RM.Pos. (reference mark position) and RM.ZC. (reference-mark zero crossovers) bar graphs. Tapping the display opens the Display options for count value screen. Status The Status display (6) shows whether there are any messages about the connected encoder.
Green: There are no messages Red: There are messages Tapping the display opens the Encoder status screen. Further information: "Encoder status screen ", Page 95 HEIDENHAIN | PWT 100, PWT 101 | Operating Instructions | 07/2024 83