heidenhain PWT 100, PWT 101 — Function settings screen

Encoder diagnostics | Diagnostics for encoders with 1 V PP /11 µA PP /1 V PP + Z1 interface 8 Deactivating the HSP function Certain encoders from HEIDENHAIN provide the HSP function that can be activated or deactivated. The HSP function is activated in the default setting. The HSP function should be deactivated only when the encoder is mounted. A corresponding message will then be shown in the PWT display screen. When an encoder is tested after it has been mounted, the HSP function should be activated. This default setting is activated when the PWT display screen is closed. For encoders with TTL interface, the HSP function is automatically set by the encoder when PWT switchover is activated.

Refer to the mounting instructions for the specific encoder. Tap More The More menu is opened Tap Functions The Display settings screen is opened Tap the Evaluation of incremental signals (HSP active) option The HSP function is deactivated Tap Back The changed settings take effect The Display settings screen is closed Evaluation options for switching signals Selection Function L1/L2 The rising and falling edges are evaluated for the L2/L/S2 switching signals. Select this setting if the encoder provides the L1 or L2 switching signals on separate signal pins. H/L (Setting for encoder standard versions) Two rising edges are evaluated for the L2/L/S2 switching signals.

This is indicated by the corresponding number- ing after the edge symbol. Select this setting if the encoder supports the Limit and Homing switching signals. H/L negated (Setting for special versions of encoders) Two falling edges are evaluated for the L2/L/S2 switch- ing signals. This is indicated by the corresponding number- ing after the edge symbol. Select this setting if the encoder supports the Limit and Homing switching signals. 90 HEIDENHAIN | PWT 100, PWT 101 | Operating Instructions | 07/2024