To Trigger the Oscilloscope EN Click on the Slope menu tab to select the desired clock edge type, which can be selected as either the rising or falling edge. 4. Data source: Click on the SDA menu tab and select CH1-CH4 or EXT as the source. Only by selecting the channel that has been connected to the signal as the trigger source can stable triggering be obtained. 5. Data Type: Click on the Data Type menu tab to set the effective code type of the data signal. It can be set to H (high level) or L (low level). 6. When: In the When menu tab, select Create Hold Trigger Condition. ⚫ Setup: When the establishment time is less than the set value, the oscilloscope triggers. ⚫ Hold:
When the hold time is less than the set value, the oscilloscope triggers. 7. Width: Click on the menu label for Width , and set the width value directly by rotating the multi-functional knob V0 or the pop-up numeric keyboard. 8.6.12 UART Trigger(Option) RS232 bus is a serial communication method used for data transmission between computers or between computers and terminals. The RS232 serial protocol transmits a character as a frame of data, with a frame structure consisting of 1 bit start bit, 5-8 bits data bit, 1 bit check bit, and 1-2 bits stop bit. The format is shown in the following figure.
This series of oscilloscopes can trigger when detecting frame start, frame end, data, verification errors, and errors of RS232 signals. Figure 8.11 Schematic Diagram of UART Protocol Click on [Trig Menu] in the button area to open the trigger menu. Copyright © Qingdao Hantek Electronic Co., LTD DPO7000 User manual 81