To Trigger the Oscilloscope EN 5. Baud Rate: Click on the Baud Rate menu tab and select the preset baud rate from the pop-up sub options. Alternatively, click again to pop up the numeric keypad and directly enter the set baud rate. 6. Data Bits: Click on the Data Bits menu tab and select the desired data width. The data width refers to the number of bits of data per frame. The data width can be selected as 5 bits, 6 bits, 7 bits, or 8 bits. 7. Data: Click on the Data menu tab and set the data directly by rotating the multi-functional knob V0 or the pop-up numeric keyboard. 8. Parity: Click on the Parity menu tab, select polarity, and you can choose None, Odd, or Even.

8.6.13 LIN Trigger (Option) This oscilloscope can be triggered on the synchronous field of LIN signals, as well as on designated identifiers, data, or frames. The format of the LIN bus data frame is shown in the following figure. Figure 8.12 Data Frame Format of the LIN Bus Click on [Trig Menu] in the button area to open the trigger menu. Copyright © Qingdao Hantek Electronic Co., LTD DPO7000 User manual 83