Hantek — DPO7000 UserManual

To Trigger the Oscilloscope EN Click on the Slope type menu tab to select the desired clock edge type. ⚫ Rising: Sampling MISO data at the rising edge of the clock. ⚫ Falling: Sampling MISO data at the falling edge of the clock. 4. Timeout: Click on the Timeout menu tab and enter the set value directly by rotating the multi-functional knob V0 or the pop-up numeric keyboard. 5. Bit X: Click on the Bit X menu tab to set the data bits that need to be operated on. 6. Data bits: Click on the Data bits key and use the pop-up numeric keypad to set the number of digits in the serial data string. The number of digits in a string can be set to any integer between 4 and 32.

8.6.16 I2C Trigger (Option) I2C (Inter Integrated Circuit Bus) integrated circuit bus, designed by NXP (formerly PHILIPS) company, is mainly used for master-slave communication between master controllers and slave devices. It is used in small data volume situations, has short transmission distance, and can only have one host at any time. I2C address addressing mode is divided into 7-bit addressing mode and 10 bit addressing mode. 7-bit addressing: In the 7-bit addressing process, the slave address starts transmission from the first byte after the start signal.

The first 7 bits of this byte are the slave address, and the 8th bit is the read write bit, where 0 represents write and 1 represents read. 10 bit addressing: The 10bit and 7bit addressing of the I2C bus are compatible, allowing devices that use both 7bit and 10bit address modes on the same bus. When transmitting a 10bit address, the first byte is a special reserved address to indicate that the current transmission is a 10bit address. Figure 8.15 Schematic Diagram of I2C Protocol Click on [Trig Menu] in the button area to open the trigger menu. DPO7000 User manua l Copyright © Qingdao Hantek Electronic Co., LTD 88