To Trigger the Oscilloscope EN 5. Sync: Click on the Sync menu tab and select the desired field or line (ScanLine, Line, Odd, Even, AllFields) to trigger the signal. ⚫ ScanLine: displays a complete line, including a portion of the previous and next lines. The oscilloscope is triggered at any line. ⚫ Line: Display a complete line, including a portion of the previous and next lines. Select a specified number of rows for the oscilloscope to trigger based on user selection. ⚫ Odd: displays multiple fields and the oscilloscope only triggers on odd fields. ⚫ Even: display multiple fields and the oscilloscope only triggers on even numbered fields. ⚫ AllFields:
Display multiple fields and trigger the oscilloscope on any field. 6. Line: Click on the Line number menu tab and set the line number in the field to be triggered. When selecting the synchronization method as the number of lines, you can specify the number of lines. The range of line count settings is related to the currently selected video standard, with settings ranging from 1 to 525 (NTSC) and 1 to 625 (PAL/SECAM). 8.6.4 Slope Trigger Slope triggering sets the positive or negative slope triggering of the oscilloscope from one level to another within a specified time.
As shown in the figure below, we define the time difference between the two points (A and B) where the high and low trigger levels intersect with the rising (falling) edge of the waveform as the positive (negative) slope time. Figure 8.5 Positive Slope Time/Negative Slope Time Click on [Trig Menu] in the button area to open the trigger menu. Copyright © Qingdao Hantek Electronic Co., LTD DPO7000 User manual 67 PAL/SECAM 25 Interlaced Scan 625