To Trigger the Oscilloscope EN 1. Type: Click on the Type menu label and select "Slope" to set the trigger. After selecting the trigger type, the current trigger setting information (including trigger type, trigger source, and trigger level) will be displayed at the top of the screen, which changes with the trigger setting. 2. Source: Click on the Source menu tab to select sources such as CH1, CH2, CH3, CH4. 3. Slope Type: Click on the Slope Type menu tab to select rising and falling edges. ⚫ Rising: Set the signal slope condition to trigger with a positive slope. ⚫ Falling: Set the signal slope condition to trigger with a negative slope. 4. When:
Click on the When menu tab and select Trigger Condition. ⚫ signal is less than the set time value can it be triggered. ⚫ >(Greater than time value): It can only be triggered when the positive or negative slope time of the input signal is greater than the set time value. ⚫ != (Not equal to time value): When the positive or negative slope time of the input signal is not equal to the set time value. ⚫ =(Equal to time value): When the positive or negative slope time of the input signal is equal to the set time value. DPO7000 User manua l Copyright © Qingdao Hantek Electronic Co., LTD 68