To Trigger the Oscilloscope EN Figure 8.7 Window Trigger ⚫ If both the high and low levels are within the waveform range, the waveform is ⚫ If the high level is within the waveform range and the low level is outside the triggered on the rising or falling edge at the same time. waveform range, the waveform will only trigger on the rising edge. ⚫ If the high level is outside the waveform range and the low level is within the Click on [Trig Menu] in the button area to open the trigger menu. waveform range, the waveform will only trigger on the falling edge. 1. Type: Click on the Type menu label and select "Window trigger" to set the trigger.
After selecting the trigger type, the current trigger setting information (including trigger type, trigger signal source, and trigger level) will be displayed at the top of the screen, which changes with the trigger setting. 2. Source: Click on the Source menu tab and select CH1-CH4. Please refer to the section on triggering data sources for details. Only by selecting the channel that has been connected to the signal as the trigger source can stable triggering be obtained. 3. Level Select: Copyright © Qingdao Hantek Electronic Co., LTD DPO7000 User manual 71