To Trigger the Oscilloscope EN Click on the menu label of Source B and select CH1-CH4 or EXT as the trigger source for Source B. Only by selecting the channel that has already been connected to the signal as the source can stable triggering be obtained. 3. Edge A and Edge B: ⚫ Edge A: Click on the Edge A menu label to select the trigger edge type of source A during delayed triggering, and you can choose either the rising or falling edge. ⚫ Edge B: Click on the Edge B menu label to select the trigger edge type of source B during delayed triggering, and you can choose either the rising or falling edge. 4. When: Select the time limit condition for delayed triggering in the When menu tab.
⚫ = [Equal to the set width value]: Only when the time difference between the set edges is equal to the set width can it be triggered (with a pulse width error of 5%). ⚫ != [Not equal to the set width value]: Only when the time difference between the set edges is not equal to the set width can it be triggered (pulse width error is 5%). ⚫ >[Greater than the set width value]: Only when the time difference between the set edges is greater than the set width can it be triggered (with a pulse width error of 5%). ⚫ edges is less than the set width can it be triggered (with a pulse width error of 5%). 5. Width:
Click on the menu label for Width , and set the width value directly by rotating the multi-functional knob V0 or the pop-up numeric keyboard. 8.6.11 Setup/Hold Trigger Establishing a hold trigger requires setting the clock channel and data channel separately. The establishment time starts from the time when the data channel crosses the power generation level and ends when the designated clock channel edge arrives; The holding time starts when the designated clock channel edge arrives and ends when the data channel crosses the touch power generation level again (as shown in the following figure).
When the establishment time or holding time is less than the preset time, the oscilloscope will trigger. Copyright © Qingdao Hantek Electronic Co., LTD DPO7000 User manual 79