12. Install tips into the heater blocks. 13. Install tip shields onto the tips. 14. Reinstall the air plenum. 15. Reinstall the head cover. 16. Reinstall the right side access cover. 18. Reinstall the top panel. See “Installing the top panel” on page 162 17. Reinstall the side panels. See “Installing the side panels” on page 164 19. Power the printer up. 20. Enter tip offset value (On tag attached to replacement toggle head): A. For printer firmware version 9.1 or newer: i. While the printer is at Idle , enter standby mode by pressing the Standby... button. ii. Enter the Service Calibration menu by pressing the following button sequence on the display panel: 3,2,3,3,1 . See Figure 148.
Figure 148: Service Calibration button sequence Button 0 Button 1 Button 2 Button 3 iii.Press the Edit Configuration button. iv. Press the Next... button. v. Press the Next... button. vi. Press the ZT2T (Z tip to tip) button. vii.Enter the calibration value from the tag that came with the replacement toggle head assembly by pressing the Increment , Decrement and Next Digit buttons. NOTE: The value for ZT2T must be between -0.0500” and 0.0500”. -218