Downloaded from https://www.everyspec.com MIL-STD-1311D 2.3 Method C (ratio method). 2.3.1 Test circuit. FIGURE 4253-5. Ratio method. 2.3.2 Procedure. With the TUT inserted in the circuit shown on figure 4253-5, portions of the input power to the TUT and of the output power from the TUT are fed to a ratio meter. The attenuator is set at a proper level. The ratio of the power to input (2) to the power at input (1) to the ratio meter is determined. The gain of the TUT is given by the formula below: Gain = dBR - dBi + dBo The ratio meter reads the ratio of the input power (2) to input power (1). Where: Gain = Gain of TUT in dB. dBR = Ratio in dB determined by the ratio meter.

dBi = Ratio in dB of the TUT input power to that portion of the TUT input power fed to the ratio meter. dBo = Ratio in dB of the TUT output power to that portion of the TUT output power fed to the ratio meter. (Sum in dB of directional coupler decoupling, attenuator, and filter (if required) losses.) 2.3.3 Precaution. 2.3.3.1 This method usually requires a square-wave modulated source for use with the ratio meter. This may produce erroneous results when used with certain kinds of pulse tubes. In the case of medium or high-power CW tubes erroneous performance measurements can result because of the reduced rf duty factor.